Galvanometer to Ammeter and Voltmeter

15 min read4 ideas5 checks1 figure to playBy Asheshanand Acharya
What you will learn 0 / 4
Sensitivity of a galvanometer

The heart of every analogue meter

A galvanometer is a coil of fine wire that can turn between the poles of a magnet. A current through the coil makes it turn, and a pointer turns with it, further for a larger current. How the magnet turns the coil belongs to the chapter Magnetic Force on Currents and the Magnetic Dipole. Here we need only two numbers: the coil's resistance \(G\), and the current \(I_g\) that swings the pointer to full scale.

How much it swings for a given current is its current sensitivity, \(\theta/I\), in divisions per milliampere. How much it swings for a given voltage across it is its voltage sensitivity, \(\theta/V = \theta/(IG)\). A galvanometer of \(50\,\Omega\) that deflects \(20\) divisions for \(1\,\text{mA}\) has a current sensitivity of \(20\,\text{div/mA}\). The \(1\,\text{mA}\) needs \(50\,\text{mV}\) across it, so its voltage sensitivity is \(20/50 = 0.4\,\text{div/mV}\).

The Class 12 practical uses the reciprocal, the figure of merit: the current per division. For the galvanometer above it is \(1/20 = 0.05\,\text{mA}\) per division. A good galvanometer has a small figure of merit and a large sensitivity.

RememberSensitivity of a galvanometer
\[ \begin{gathered} \text{current sensitivity} = \dfrac{\theta}{I} \\ \text{voltage sensitivity} = \dfrac{\theta}{IG} \end{gathered} \]
Holds whena galvanometer whose deflection is proportional to the current through its coil
Symbols\(\theta\) deflection, \(I\) current through the coil, \(G\) resistance of the coil
Unitsdivisions per mA and divisions per mV

Why can the bare coil not measure amperes or volts? Its wire is fine, and \(I_g\) is often a milliampere or less. A current of a few amperes would drive the pointer hard against its stop and burn the coil. Full scale needs only \(I_gG\) across it, \(0.1\,\text{V}\) for \(1\,\text{mA}\) and \(100\,\Omega\). So a few volts across it would wreck it too.

ExampleTwo sensitivities

A galvanometer of resistance \(100\,\Omega\) reads full scale, \(30\) divisions, at \(1\,\text{mA}\). Find its current sensitivity, its voltage sensitivity and its figure of merit.

Try it on paper first. The solution is one tap away when you are ready.
The shunt makes an ammeter

A shunt makes an ammeter

An ammeter must carry the whole current it measures, so it goes in series with the branch. To read \(1\,\text{A}\) with a coil that takes only \(1\,\text{mA}\), give the rest of the current another road. A small resistance, the shunt \(S\), is joined in parallel with the coil. Most of the current takes the shunt, and a fixed small share passes through the coil.

The coil and the shunt are in parallel, so they have the same voltage across them. At full scale the coil carries \(I_g\) and the shunt the rest, \(I - I_g\). So \(I_gG = (I - I_g)S\), and the shunt must be \(S = I_gG/(I - I_g)\).

RememberShunt for an ammeter
\[ \begin{gathered} I_g G = (I - I_g)S \\ S = \dfrac{I_g G}{I - I_g} \end{gathered} \]
Holds whena galvanometer of resistance \(G\) and full-scale current \(I_g\), to read \(I\) at full scale
Symbols\(S\) shunt, \(G\) coil resistance, \(I_g\) full-scale current of the coil, \(I\) full-scale current of the ammeter
Units\(S\) and \(G\) in Ω, currents in A
A galvanometer, shunted and multipliedA galvanometer of resistance 100 ohms that reads full scale at 1 milliampere, between two terminals. The chips turn it into three instruments. As a bare galvanometer it reads up to 1 milliampere. As an ammeter of range 1 ampere it has a shunt of 0.1 ohm in parallel: at 0.5 ampere the coil carries 0.5 milliampere and the shunt 499.5 milliamperes, so the shunt takes 99.9 percent of the current. As a voltmeter of range 10 volts it has 9900 ohms in series: at 5 volts the coil carries 0.5 milliampere and the series resistance drops 4.95 volts, 99 percent of the voltage. A slider sets the measured quantity as a fraction of the range, and in every case the needle stands at that same fraction of full scale. G Ig I = 0.50 mA through the coil full scale: 1 mA, or 0.1 V across G
SimA galvanometer of \(100\,\Omega\) that reads full scale at \(1\,\text{mA}\). The chips make it a \(1\,\text{A}\) ammeter or a \(10\,\text{V}\) voltmeter. The slider sets what is measured, as a fraction of the range.

For \(G = 100\,\Omega\) and \(I_g = 1\,\text{mA}\), a \(1\,\text{A}\) range needs \(S = 0.1/0.999 \approx 0.1\,\Omega\). In the figure above the shunt carries \(999\,\text{mA}\) of every ampere, \(99.9\,\%\), and the coil \(0.1\,\%\). The ammeter's own resistance is \(GS/(G + S) = 0.1\,\Omega\). A \(5\,\text{A}\) range needs \(0.1/4.999 \approx 0.02\,\Omega\): a larger range needs a smaller shunt.

The split is fixed by the ratio of the resistances, so the coil always carries the same fraction \(S/(S + G)\) of the current. Half an ampere through the \(1\,\text{A}\) ammeter sends \(0.5\,\text{mA}\) through the coil, and the pointer stands at half scale. The scale stays linear. The shunt trades sensitivity for range: if full scale is \(20\) divisions, one division now means \(1/20\,\text{A} = 50\,\text{mA}\).

When \(S\) is much smaller than \(G\), the ammeter's resistance \(GS/(G + S) = S/(1 + S/G)\) is almost exactly \(S\). A shunt of \(1\,\Omega\) on a \(100\,\Omega\) coil lets \(1/101\) of the current, about \(1\,\%\), through the coil. A second \(0.1\,\Omega\) joined across a \(0.1\,\Omega\) ammeter takes half of every current, so the meter's full scale becomes \(2\,\text{A}\).

ExampleShunting a galvanometer

A galvanometer of resistance \(50\,\Omega\) reads full scale at \(2\,\text{mA}\). Find the shunt that makes it an ammeter of range \(2\,\text{A}\), and the resistance of that ammeter.

Reveal one step at a time and finish the rest yourself.
Checkpoint · numericUnanswered

A galvanometer of resistance \(40\,\Omega\) reads full scale at \(5\,\text{mA}\). What shunt makes it an ammeter of range \(2\,\text{A}\), in ohms?

Ω± 2 %
The series resistance makes a voltmeter

A series resistance makes a voltmeter

A voltmeter must have the voltage it measures across it, so it goes in parallel with the element. To read \(10\,\text{V}\) with a coil that needs only \(0.1\,\text{V}\), a large resistance, the multiplier, is joined in series with the coil. At full scale the current through both is \(I_g\), so \(V = I_g(G + R)\) and \(R = V/I_g - G\).

RememberSeries resistance for a voltmeter
\[ \begin{gathered} V = I_g (G + R) \\ R = \dfrac{V}{I_g} - G \end{gathered} \]
Holds whena galvanometer of resistance \(G\) and full-scale current \(I_g\), to read \(V\) at full scale
Symbols\(R\) series resistance, \(G\) coil resistance, \(I_g\) full-scale current, \(V\) full-scale voltage of the voltmeter
Unitsresistances in Ω, \(V\) in V, \(I_g\) in A

For \(G = 100\,\Omega\) and \(I_g = 1\,\text{mA}\), a \(10\,\text{V}\) range needs \(10/0.001 - 100 = 9900\,\Omega\). The voltmeter's resistance is \(10\,000\,\Omega\). Take the voltmeter chip in the figure above: at \(5\,\text{V}\) the coil carries \(0.5\,\text{mA}\), and the \(9900\,\Omega\) drops \(4.95\,\text{V}\), \(99\,\%\) of the voltage. The same galvanometer with \(9900\,\Omega\) across a \(5\,\text{V}\) cell therefore stands at half scale.

A galvanometer of \(50\,\Omega\) and \(2\,\text{mA}\) needs \(5/0.002 - 50 = 2450\,\Omega\) for a \(5\,\text{V}\) range. For \(20\,\text{V}\) it needs \(9950\,\Omega\), so \(7500\,\Omega\) more must be added in series. A meter with two ranges uses two resistances in a chain. With the \(100\,\Omega\), \(1\,\text{mA}\) coil, \(R_1 = 9900\,\Omega\) gives \(10\,\text{V}\). Adding \(R_2 = 100/0.001 - 100 - 9900\) \(= 90\,000\,\Omega\) gives \(100\,\text{V}\).

Checkpoint · numericUnanswered

A galvanometer of resistance \(40\,\Omega\) reads full scale at \(5\,\text{mA}\). What resistance in series makes it a voltmeter of range \(15\,\text{V}\), in ohms?

Ω± 2 %
CheckpointUnanswered

Which statement about ideal meters is correct?

How a meter disturbs the circuit

Every meter changes the circuit

A meter becomes part of the circuit it measures. An ammeter in series adds its own resistance. A \(12\,\text{V}\) cell of negligible internal resistance drives \(12/5 = 2.4\,\text{A}\) through \(5\,\Omega\). Insert an ammeter of \(1\,\Omega\), and the current falls to \(12/6 = 2\,\text{A}\). The meter reads \(2\,\text{A}\), \(16.7\,\%\) below the current that flowed without it.

A voltmeter in parallel draws current. Two \(1000\,\Omega\) resistors in series on \(12\,\text{V}\) each have \(6\,\text{V}\) across them. Put a \(1000\,\Omega\) voltmeter across one. The pair becomes \(500\,\Omega\), the current rises to \(12/1500 = 8\,\text{mA}\), and the meter reads \(8\,\text{mA} \times 500\,\Omega = 4\,\text{V}\). It has changed the voltage it came to measure.

A voltmeter that loads the circuitA 12 volt cell of negligible internal resistance drives two 1000 ohm resistors in series, and a voltmeter is joined across the lower one. The true voltage across it is 6 volts. First chip, an ideal voltmeter of infinite resistance: it draws no current and reads 6 volts. Second chip, a 1000 ohm voltmeter: in parallel with the resistor it makes 500 ohms, the current rises to 8 milliamperes and the reading falls to 4 volts. Third chip, a 10 kilohm voltmeter, a 1000 ohm per volt meter on its 10 volt range: the parallel pair is 909 ohms and the reading is 5.7 volts, much closer to the truth. 1000 Ω 1000 Ω 12 V V ideal reads 6.00 V, true value 6 V
Fig 6.1Two \(1000\,\Omega\) resistors on \(12\,\text{V}\), with a voltmeter across the lower one. The chips change the voltmeter's resistance.

In Fig 6.1 a meter rated \(1000\,\Omega\) per volt, used on its \(10\,\text{V}\) range, has \(10\,\text{k}\Omega\). Across the lower resistor it makes \(909\,\Omega\) and reads \(5.7\,\text{V}\), much nearer the truth. So an ammeter should have as little resistance as possible, and a voltmeter as much as possible. An ideal meter has zero resistance as an ammeter and infinite resistance as a voltmeter.

Property
Ammeter
Voltmeter
Joined
in series with the branch
in parallel with the element
Made from a galvanometer by adding
a small shunt \(S\) in parallel
a large resistance \(R\) in series
Design formula
\(S = I_gG/(I - I_g)\)
\(R = V/I_g - G\)
Its own resistance
very small, about \(S\)
very large, \(V/I_g\)
Ideal value
zero
infinite
Checkpoint · numericUnanswered

Two \(5000\,\Omega\) resistors are joined in series across a \(10\,\text{V}\) supply of negligible internal resistance. A voltmeter of resistance \(5000\,\Omega\) is joined across one of them. What does it read, in volts?

V± 2 %

An ammeter must never be joined straight across a cell. Its resistance is a fraction of an ohm, so the current is limited only by the cell's internal resistance. A \(1.5\,\text{V}\) cell of \(0.5\,\Omega\) would push \(1.5/0.6 = 2.5\,\text{A}\) through a \(0.1\,\Omega\) ammeter of \(1\,\text{A}\) range, far past its full scale. The shunt and the coil overheat and may burn.

Checkpoint · in your own wordsUnanswered

A student joins an ammeter directly across the terminals of a cell. Why is this dangerous for the meter?

In the world
The range switch of a multimeter

An analogue multimeter holds one galvanometer and a rotary switch. On a current range the switch joins one of several shunts across the coil; the smaller the shunt, the larger the range. On a voltage range it puts one of several series resistances in the chain. A meter rated \(20\,\text{k}\Omega\) per volt, a common rating, has a coil of \(50\,\mu\text{A}\) full scale. With a coil of \(100\,\Omega\) its \(10\,\text{V}\) range needs about \(200\,\text{k}\Omega\) in series, and a \(1\,\text{A}\) range needs a shunt of only about \(5\,\text{m}\Omega\).

Story · Jacques-Arsène d'Arsonval, 1882
A coil that swings, not a needle

Early galvanometers used a magnetised needle turned by a current in a coil around it, and nearby iron or the Earth's field upset them. In 1882 the French physician and physicist Jacques-Arsène d'Arsonval, working with Marcel Deprez, turned the design round. A light coil hung between the poles of a fixed permanent magnet and turned itself. The moving-coil galvanometer was steady and sensitive, and every analogue ammeter and voltmeter since has been built round it.

Exam traps
  1. Joining the shunt in series with the coil.
    The shunt must share the coil's voltage and take most of the current, so it goes in parallel.
  2. Forgetting to subtract \(G\) when finding the series resistance.
    \(V/I_g\) is the whole voltmeter's resistance. The added resistance is \(V/I_g - G\).
  3. Thinking that a real meter reads the value that was there before it was joined.
    An ammeter lowers the current by adding resistance; a voltmeter lowers the voltage by drawing current.
  4. Using an ammeter to check whether a cell is alive.
    Across a cell an ammeter nearly shorts it and burns. Use a voltmeter, which draws almost nothing.